Automatic Titrators Orion Star™ T920


For redox titrations by equivalence point method or preset mV endpoint, e.g. sulfite/SO2, reducing sugar, ascorbic acid, peroxide as well as dissolved oxygen and dissolved organic substance.

Scope of supply:
START9200: Titrator incl. 20 ml burette, stirrer probe, dispenser probe, standard tubing kit, 1 l plastic bottle, GL38 bottle cap with drying tube, computer cable, manual on USB drive, 110-240 V power supply
START9201: As START9200, additionally with 9778BNWP redox glass electrode
Specifications
Titration precision:±0.5 % RSD
Speed of stirrer probe:250 ... 3700 rpm, 5 selectable speeds
mV range:
Resolution:
Rel. accuracy:
-2000.0 ... +2000.0 mV
0.1 mV
±0.2 mV
Temperature range:
Resolution:
Rel. accuracy:
-5,0 ... 100.0 °C
0.1 °C
±0.2 °C
Display:5.7'' Touchscreen, 640 x 480 resolution
Burette sizes:10 ml, 20 ml, 50 ml
Burette resolution:25600 micro-steps per motor revolution
2 million micro-steps over full stroke range
Dosing accuracy:Acc. to ISO 8655-3
Dimensions (W x D x H):406 x 254 x 356 mm
Weight:5.67 kg
Power supply:100 ... 240 V, 50/60 Hz
Certification:TÜV 3-in-1, FCC, EN/EIC 61326-1, IEC 61010, IP-51
 
TypePKCat. No. Ref. Fabricante
START920014672085 START9200
START920114672086 START9201
Accesorios:
Accesorios para Titradores Automticos Orion Star™ / Thermo Elect.LED
Alternativas:
Titradores Automáticos Orion Star™ T910 / Thermo Elect.LED
 Automatic Titrators Orion Star™ T910 / Thermo Elect.LED
 Automatic Titrators Orion Star™ T910 / Thermo Elect.LED